beautypg.com

Posey Soft Split Side Rail Protectors User Manual

Posey, Split side rail protectors, Monitoring

background image

1

2

$

%

Fig. 2

Fig. 3

Posey Split Side Rail Protectors feature a universal design with

multiple versatile hook and loop straps for easy attachment to

many beds and side rails. The soft, padded design helps protect

against hard side rail surfaces and is covered in a water repellent

vinyl that can be wiped clean between uses. Provides a visual

deterrent against patients exiting the bed between the side rails and

depending on the bed system, these protectors may help mitigate

and safeguard against FDA/HBSW entrapment zones 1 and 5

1

. Meets

CA #117, Section E Flame Retardancy Standards.

REF

5705

For older, open bar side rail designs,

36”L x 16”H x ½”thick (91 cm x 41 cm x 1 cm)

REF

5705LS For newer, encapsulated side rail designs,

36”L x 16”H x ½” thick (91 cm x 41 cm x 1 cm)

(Two more upper straps than model 5705. All four

upper straps are longer than model 5705)

Monitoring

Posey Split Side Rail Protectors will not prevent a

determined patient from exiting the bed, going over or attempting

to go between or through side rail gaps. DO NOT articulate bed if

product is applied.

ALwAyS follow your facility’s policies and

procedures for patient assessment, monitoring and rehabilitation.

Application Instructions:

1. Inspect product prior to use. DO NOT use soiled or damaged

products.

2. Place the pad inside and between the upper and lower side rails of

the bed with the attachment straps pointing outward (Fig. 1).

3. Secure the bottom attachment straps to the bed frame. At least

two of the three straps need to be secured for proper use.

4. Position and attach the remaining straps to the bed side rails by

(Fig. 2):

A. Wrapping the loop closure completely around the attaching

point;

B. Pulling the hook material over the loop material;
C. Excess strap length may be cut if desired.
5. While not all straps may be applicable for use on some bed

models, ensure that enough of the straps are attached to secure

the pad on the bed (Example of a properly installed 5705 shown

in Fig. 3).

6. Check for proper fit. Modified beds may not be compatible with

Posey Split Side Rail Protectors.

Cleaning Instructions:

Wipe clean with soap and water or disinfectant.

DO NOT use cleaners that contain Phenol or Benzyl.

A

Disposal

Properly dispose of the product

per facility’s policy for BIOHAZARDOUS

materials.

Storage And Handling

This device is designed for use in normal indoor

environments. This device may be stored in

ambient warehouse temperatures at normal

humidity levels. Avoid excess moisture or high

humidity that may damage product materials.

$

%

Fig. 1

C

B

5705LS

© 2010 Posey Company. All rights reserved.

M1628 REV B 071810

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

1

Guidance for Industry and FDA Staff - Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment, March 10, 2006.

Posey

®

Split Side Rail Protectors

Bed SAFety