Posey, Self-releasing chair belt sensors (cont.), For model 8372l (hook and loop sensor belt, long) – Posey Mobile Chair Belt Sensor User Manual
Page 2: Cleaning

For Model 8358, 8371 and 8372:
1. Draw strap ends down under chair seat at
a 45˚ angle (fig. 3).
2. Choose a site to attach straps at a juncture
of the frame behind and under the seat,
out of the patient’s reach. Make sure straps
cannot slide in any direction, or change
angle or position (fig. 4).
3. Wrap each strap around its attachment point. Thread tail of each
strap through both metal rings, then back over one ring and through
the other metal ring. Pull straps until secure.
4. Position patient as far back on seat as possible, with buttocks
against the back of the chair.
5. Draw straps across patient’s hips at a 45° angle and fasten belt.
Mobile Chair Belt and EZ Clean Alarm Belt.
• Insert male end into buckle until it clicks shut. Pull firmly on
straps to make sure buckle is locked.
• Adjust belt by pulling on loose strap.
• To unfasten, press orange button and remove male end.
Hook and Loop Belt.
Mate hook and loop straps for snug fit. Pull YELLOW strap to
unfasten.
For Model 8372L (Hook and Loop Sensor Belt, Long)
1. Draw straps down over patient’s lap at a
45˚ angle (fig. 5).
2. Position strap ends behind chair back as
low as possible. Thread tail of one strap
through both metal rings of other strap
(fig. 6), then back over one ring and
through the other metal ring (fig. 7).
3. Pull straps until secure (fig. 8).
4. Mate hook and loop straps for snug fit.
Pull YELLOW strap to unfasten.
Adjust Straps:
ALWAYS adjust straps so belt is snug, but
does not interfere with breathing. You should
be able to slide your open hand (flat)
between the belt and the patient.
If Hook and Loop straps
become wet, alarm may not sound. Excess
moisture may prevent sensor from activating,
increasing risk of injury. If straps become
wet, discontinue use until completely dry.
Cleaning:
1. Wipe clean with a cloth dampened with disinfectant. Take extra
care with sensor plug.
2. Dry hook and loop straps completely before use. Use a stiff brush
to remove dust and lint from hook and loop.
3. NEVER:
• Use any cleaner that contains Phenol or Benzyl;
• Immerse in liquid; or
• Sterilize with heat.
4. Dry with a clean cloth.
Storage and Handling:
This device is designed for use in normal indoor environments, This
device may be stored in ambient warehouse temperatures at normal
humidity levels. Avoid excess moisture or high humidity that may
damage product materials.
Disposal:
Properly dispose of the product per facility policy
for BIOHAZARDOUS material.
Limited Warranty:
This product is warranted for 1 year from date of first use. See Posey
Catalog for details.
Fig. 4
Mobile Chair Belt/Hook and Loop Alarm Belt/EZ Clean
Alarm Belt rear attachment
8372L
Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014 • Web: www.posey.com
To reduce the risk of severe injury or death:
• Check that alarm is ON and working properly EACH TImE
before leaving patient unattended. Press self-release button to
unfasten buckle, or separate hook and loop straps. Alarm
should activate each time you do this. STOP USE AT ONCE if
alarm fails any test.
• DO NOT leave patient unattended unless you are sure patient
understands purpose for and how to remove the belt. Check
that belt is snug, but does not interfere with breathing.
• NEVER jerk or pull on sensor cord to remove plug from alarm.
Doing so will damage the cord or plug and may cause sensor to
fail. ALWAYS use plastic tab to release plug.
45˚
Fig. 3
45˚
Fig. 5
Fig. 7
Fig. 8
Fig. 6
Posey
®