Record a message, Alarm mode selection – Posey Sitter Elite® User Manual
Page 2
![background image](/manuals/408217/2/background.png)
Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com • Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014
7. Tone Selection
8. Volume Selection
9. Setting a Time Delay
Press the TONE button until the tone
that you would like to use plays. Each
time you press the TONE button, the
selection changes and a two (2) second
sample of the tone is played. There are
5 tones to choose from. The last sample
played is the tone selected for use.
Press the VOLUME button to scroll
through the volume selections. Each
time you press the VOLUME button, the
selection changes and a two (2) second
sample of the volume is played. Continue
pressing the VOLUME button until you
hear the desired volume setting. The
last sample heard will be the volume
selected for use.
Move the delay switch to 0, 1 or 2
to change the time delay.
10. Record a Message
The green light for Voice and Tone, Tone, Voice or the
red light for HOLD/SUSPEND must be flashing. Press
and hold down the RECORD button. Wait for the beep
before you begin speaking. Speak in a normal voice into
the microphone aperture labeled “MIC” on the top front
of the case. The recording will stop when you release the
button or when you have exceeded 10 seconds. You will
hear a second beep if you have exceeded the 10 second
time limit. This recorded message will play whenever a
voice mode is selected. The louder you speak into the
microphone, the louder the voice playback.
6. Alarm Mode Selection
Turn the alarm on.
Connect a sensor
with weight
applied. Open the
selection door to
display the option
buttons.
Turn the alarm on.
Connect a sensor
with weight
applied. Open the
selection door to
display the option
buttons.
Turn the alarm on.
Connect a sensor
with weight
applied. Open the
selection door to
display the option
buttons.
Turn the alarm
on. Open the
selection door
to display
the option
buttons.
Turn the alarm
on. Connect
a sensor with
weight applied.
Open the
selection door
to display the
option buttons.
Press the MODE selection button
until the LED light flashes green
above the mode you would like
to use.
TONE
TONE
0
1
2
TONE
0
1
2
TONE
0
1
2
TONE
0
1
2
TONE
TONE
TONE
TONE
MODE
BUTTON
TONE
BUTTON
VOLUME
BUTTON
DELAY
SWITCH
TONE
0
1
2
RECORD
BUTTON
After set up, ALWAYS test the alarm with the sensor before leaving the patient unattended. DO NOT use any alarm that does not
sound each time it is activated.