Note…

When the pen is switched from

Measurement Mode to Calibration Mode,

it will display the factory default value.

Therefore, if the pen was previously

calibrated, the display may appear to jump

to the factory default value when entering

Calibration Mode

MEASUREMENT PROCEDURES
1. Remove the probe cover from the pen.

4

Remove the cover

from the bottom

of meter to expose

the electrode.

Electrode

3. The meter is default in auto-ranging

status once powered on.

Auto-ranging determines and selects

a range which gives you the greatest

resolution and accuracy. Alternatively,

you can manually select the ranges.

For example, if you prefer the meter

to display a reading as 0.50 mS

instead of 500 µS, you just need to

press “ “ m

select 19.99 mS range when you

are in normal measurement mode.

ore than 2 seconds

to

NOTE:

Accuracy is a percent of full-scale,

so using your meter in the lowest

range will result in the greatest

accuracy.

The LCD will display E02/E03 if the

measured value is below or over the

limit value, please re-select the range.

1.

2.

See page 12

for the detail

Remove the cover from

the bottom of pen to

expose the electrode.

4

Remove the cover

from the bottom

of meter to expose

the electrode.

Electrode

3. The meter is default in auto-ranging

status once powered on.

Auto-ranging determines and selects

a range which gives you the greatest

resolution and accuracy. Alternatively,

you can manually select the ranges.

For example, if you prefer the meter

to display a reading as 0.50 mS

instead of 500 µS, you just need to

press “ “ m

select 19.99 mS range when you

are in normal measurement mode.

ore than 2 seconds

to

NOTE:

Accuracy is a percent of full-scale,

so using your meter in the lowest

range will result in the greatest

accuracy.

The LCD will display E02/E03 if the

measured value is below or over the

limit value, please re-select the range.

1.

2.

See page 12

for the detail