9. When “CAL” stops flashing on the

LCD, press

SET for less than 1 second

to confirm the value. The pen will

return to Measurement Mode for salt

measurement. To exit Calibration Mode

without saving, press

SET for more than

2 seconds. This allows you to retain the

pen’s previous calibration data for the

current range as you proceed. If “CAL”

flashes on the LCD continually, check

that the calibration solution is stable and

that the input value (Step 8) is correct.

10. Repeat Steps 1-9 for other ranges as

needed.

Note…

If the standard value is over the measuring

range or 10% less, the displayed value will

be equal to the range limit or 10% of the

range limit.