Sper Scientific 850037 Large Display Conductivity Pen User Manual

Page 12

background image

12

To manually select the measurement range:

1. Press

SET to turn the pen on.

2. Press ▲ for more than 2 seconds to

select range 1. Press ▲ for more than

2 seconds to select range 2.

CALIBRATION
Selecting the Calibration

Standard Solution
For best results, select a NaCl standard

near the sample value that you are

measuring. Alternatively, use a calibration

solution value that is approximately ⅔ of

the full scale of the measurement range

that you plan to utilize.

Do NOT reuse the calibration solution.

Contaminants in the solution will affect the

calibration and the accuracy. Use fresh

solution each time.

The previous calibration data will be

replaced after re-calibrating. However,