Salinity Calibration

1. Place the probe into deionized or

distilled water for about 30 minutes to

rinse the probe.

2. Select the NaCl standard for calibration.

3. Pour 3 cm (deep) of solution into two

separate containers (A & B).

4. Press

SET to turn the pen on.

5. Rinse the probe in one of the

containers. Gently stir the probe.

6. Dip the rinsed probe into the other

container. Tap the probe on the bottom

of the container to remove air bubbles.

Let the probe stabilize to the solution

temperature (allow 15 minutes).

7. Press

CAL for more than 2 seconds to

begin calibration. “CAL” and the salinity

value of the solution will flash on the LCD.

8. Press ▲ or ▼ to adjust the value on the

primary display to match the value of

the standard buffer. You can adjust the

salinity reading ± 30% from the detected

value. If the detected value and the

standard value differ by more than ±

30%, clean or replace the pen.