6.2 Troubleshooting Chart

Problem

Possible Cause

Solution

Unit does not turn on,

Wall switch power is off.

Turn on wall switch power.

and AC adapter is
properly connected to

Defective AC adapter.

Replace AC adapter.

a wall outlet.

AC adapter is not plugged into

Check connection from the

LifeStyle.

AC adapter to LifeStyle.

General malfunction

Compressor failure.

Replace compressor.

alarm is activated.

Battery lasts less

Battery not fully charged.

Ensure that battery

than 50 minutes.

charges for at least 2 to
2 ½ hours.

Battery failure.

Replace battery.

Oxygen concentration is

Compressor has reached its Replace compressor.

below specifications in expected life (~3,000 hrs).

test mode.

Unit does not deliver

Cannula kinked or not

Check cannula and the

oxygen.

connected.

cannula connection.

LifeStyle is not turned on.

Turn LifeStyle on to the
prescribed setting.

Battery is dead.

Replace battery.

6-2