6.2 Troubleshooting Chart
Problem
Possible Cause
Solution
Unit does not turn on,
Wall switch power is off.
Turn on wall switch power.
and AC adapter is
properly connected to
Defective AC adapter.
Replace AC adapter.
a wall outlet.
AC adapter is not plugged into
Check connection from the
LifeStyle.
AC adapter to LifeStyle.
General malfunction
Compressor failure.
Replace compressor.
alarm is activated.
Battery lasts less
Battery not fully charged.
Ensure that battery
than 50 minutes.
charges for at least 2 to
2 ½ hours.
Battery failure.
Replace battery.
Oxygen concentration is
Compressor has reached its Replace compressor.
below specifications in expected life (~3,000 hrs).
test mode.
Unit does not deliver
Cannula kinked or not
Check cannula and the
oxygen.
connected.
cannula connection.
LifeStyle is not turned on.
Turn LifeStyle on to the
prescribed setting.
Battery is dead.
Replace battery.
6-2
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)