beautypg.com

AirSep MN110-1 User Manual

Page 19

background image

4-3

A list of undesirable chemical agents includes, but is not limited to, the following,
according to the plastic manufacturer: alcohol and alcohol-based products (isopropyl
alcohol), concentrated chlorine-based products (ethylene chloride), and oil-based
products (Pine-Sol, Lestoil). These are not to be used to clean the plastic housing of
LifeStyle. These chemicals and the cleaners that contain them will crystallize the plastic
and break down the plastic’s molecular properties.

Clean the air intake / gross particle filter with warm soapy water between each patient’s
use. The patients should be instructed to clean this filter at least once per week, de-
pending on the environment, during normal operation.

Change the felt filter after 3,000 hours of use.

Change the product filter after 3,000 hours of use. It is not necessary to change this

filter between patients even if the previous patient had a communicable disease or
infection.

Allow LifeStyle to air dry, and then retest oxygen concentration before you return the
unit to your inventory.