6 lifestyle classifications – AirSep MN110-1 User Manual
Page 10

1.6
LifeStyle Classifications
Type of protection against electric
Protection from electric shock is achieved by
shock:
double insulation.
Class II
Degree of protection against
Equipment providing a particular degree of
electric shock:
protection against electric shock regarding
Type BF
1) allowable leakage current
2) reliability of protective earth connection
(if present).
Not intended for direct cardiac application.
Independent testing for
Tested by UltraTech Group of Labs to be in
Medical Electrical Equipment
compliance with IEC 60601-1:1988 +
Standard
A1:1991, A2:1995
Protection against potential
Tested by UltraTech Group of Labs to be in
electromagnetic or other
compliance with IEC 601-1-2 (EMC)
interference between the
equipment and other devices:
Method of cleaning and
Please refer to “Provider Maintenance,”
infection control allowed:
Section 4.0 of this LifeStyle Service Manual.
Degree of safety of
Equipment not suitable for such application.
application in the presence
of flammable anesthetic
gases:
Mode of operation:
Continuous duty
FAA SFAR 106:
RTCA DO160D, Section 21,
Category M-Compliant
1-6