3 usage at night or with extended tubing – AirSep MN110-1 User Manual
Page 16
3-2
2.
Remove the filter from the unit and wash it in a solution of warm water
and soap.
3.
Rinse the filter thoroughly, and remove excess water with a soft,
absorbent towel. Ensure that the filter is dry before replacing it.
4.
Replace the dry filter.
3.2.2 Checking the Hour Meter
The compressor, product filter, and felt filter must be replaced after 3,000
hours of unit use. Instruct the patient to take note of the reading on the hour
meter, located beneath the air intake gross particle filter, whenever cleaning
the air intake / gross particle filter. When the reading approaches 3,000
hours, instruct the patient that it is important to notify the Equipment Provider.
3.3
Usage at Night or with Extended Tubing
Instruct patients that AirSep recommends that LifeStyle should be used while sleeping
or with extended tubing (up to 25 ft, 7.5 m) only under the direction of their clinician.
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)