beautypg.com

0 introduction – AirSep MN171-1 User Manual

Page 5

background image

MN171-1 Rev. A 02/14

1.0 Introduction

1.1 Equipment Provider Responsibility

All Equipment Providers of the Focus

Portable Oxygen Concentrator

must assume responsibilities for handling, operational check-out, patient

instruction, and maintenance. These responsibilities are outlined below

and throughout this manual.


This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to
communicate discomfort while using this unit may require additional monitoring. Patients with
hearing and/or sight impairment(s) may need assistance with monitoring alarms.


As an Equipment Provider, you must do all of the following:


Inspect the condition of each Focus unit immediately upon

delivery to your business location. Note any sign of damage,

external or internal, on the delivery receipt, and report it

directly to both the freight company and AirSep Corporation

immediately.


Check the operation of each Focus unit before delivery to a

patient. Always operate the unit for a

minimum of 10 minutes,

and check that the oxygen concentration level is within

specifications, as described in Section 5 of this manual.


Deliver Focus units only to patients authorized by a physician’s

prescription. Focus must not be used as a life-supporting

device.


Instruct patients how to use Focus in conjunction with the

Patient Manual.


Instruct patients to notify their physicians and/or Equipment

Providers if they experience any signs of discomfort.


Be available to service each patient at any time.


Maintain Focus in accordance with Section 7.


Repair components and replace parts only as outlined in this

manual. Use only AirSep parts for replacement in Focus Oxygen

Concentrators.