beautypg.com

Recommended operating environments – AirSep 15062781 User Manual

Page 9

background image

Companion 5 User Manual — 9

Recommended Operating Environments

For proper use of the device, the following chart provides

important information concerning the recommended operating

environments, or operating conditions.

specifications

Operating

Temperature

41º F to 104º F (5° to 40°C)

Operating

Humidity

10% - 95% at an 82.4° F (28° C) dew point

Transport/

storage

Temperature

-40º F to 158º F (-40º C to 70º C)
Humidity: 10% - 95% Non-Condensing

electrical

Use no extension cords. Use no electrical outlets con-
trolled by a switch.

Operating

Altitude

-1253–9879 feet (-382–3011 meters)

Placement

DO NOT block the air inlet (top rear) or the exhaust vent
(back bottom left and right). Place device a minimum of
12 inches (30 cm) away from walls, draperies, furniture,
etc.

environment Must be smoke, pollutant, and fume free.

Operating

Time

up to 24-hours a day

Note: If the Oxygen Concentrator has been stored for an extended period
of time outside its normal operating temperature range, the unit should
be allowed to return to normal operating temperature before being turned
on. (Refer to the Specifications section in this manual.)

Proper Placement of the Companion 5

Select a location for the device that avoids the intake of smoke,

fumes and pollutants. Correct placement of the device should

allow intake of air through the air inlet at the top rear of the

cabinet and allow exhaust air to freely leave the exhaust vent at

the bottom left of the device.

Air inlets are located on the upper back and bottom of the

concentrator. Locate the unit so that there is at least 12 inches (30

cm) of space between the concentrator and any walls, furniture,

curtains, or other obstructions.

Place the device such that the alarms can be heard.

Position the oxygen supply tubing in such a way that it does not

kink or occlude.

WARNING: THIs devICe sUPPlIes HIGH-CONCeNTRATION
OXYGeN THAT PROmOTes RAPId BURNING. dO NOT AllOW
smOKING OR OPeN FlAmes WITHIN THe sAme ROOm

OF (1) THIs devICe, OR (2) ANY OXYGeN-CARRYING ACCessORY.
FAIlURe TO OBseRve THIs WARNING CAN ResUlT IN seveRe FIRe,
PROPeRTY dAmAGe, ANd/OR CAUse PHYsICAl INJURY OR deATH.

CAUTION: do not operate unit in a restricted or confined space (i.e.,
a small case) where ventilation can be limited. This can cause the
Oxygen Concentrator to overheat and impair performance. do not
allow either the air intake or the air outlet vents to become blocked.
This can cause the Oxygen Concentrator to overheat and impair
performance.