AirSep 15062781 User Manual
Page 6
6 — 15062781 B
Use the Flow Control Knob to select the flow prescribed by your
physician.
WARNING: IT Is veRY ImPORTANT TO seleCT ONlY THe
PResCRIBed level OF OXYGeN. dO NOT CHANGe THe
FlOW seleCTION UNless YOU HAve BeeN dIReCTed TO dO
sO BY A lICeNsed ClINICIAN. THe OXYGeN CONCeNTRATOR mAY Be
Used dURING sleeP UNdeR THe ReCOmmeNdATION OF A qUAlI-
FIed ClINICIAN.
Breathe normally through the nasal cannula.
To adjust flow rate:
Turn counter-clockwise to increase flow.
Turn clockwise to decrease flow.
The middle of the ball indicates flow rate.
There are two sets of lines (front and back). To properly read the
flow meter, adjust your viewing angle so front and back lines line
up exactly. See picture below.
5
Adjust Flow setting to Prescribed level
Plug the AC power cord into an AC outlet.
Toggle the Power Switch to the On position. When the
Companion 5 is plugged in properly and the switch is toggled
into the “ON” (I) position, the external power indicator will
appear on the user control panel.
When the Companion 5 is plugged in properly, a green indicator
on the LED Display will light up. After initially powering ON the
device, please allow up to ten (10) minutes for the device to reach
its performance specifications.
Note: All LED lights will illuminate upon start-up. After the concentrator
completes the warm-up cycle, only the green light will remain on.
For OCSI Models: The O2 Light will illuminate for four (4)
seconds upon start-up. After this initial start-up, this light
is disabled for approximately five (5) minutes while the
concentration rises to specifications.
4
Power On and Warm Up
Power
Indicator
LED
Power
Switch