beautypg.com

AirSep MN165-1 G User Manual

Page 24

background image

[Read the Important Safety Rules section before operating this equipment.]

21

MN165-1 Rev G

Recommended separation distances between

portable and mobile RF communications equipment and the Unit

The unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the unit
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the unit as recommended below, according to the
maximum output power of the communications equipment.

Rated maximum output

power of transmitter

W

Separation distance according to frequency of transmitter
(m)

from 150kHz to 80MHz

d= 1.2 x

√P

from 80MHz to 800MHz

d= 1.2 x

√P

from 800MHz to 2.5GHz

d= 2.3 x

√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the
transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.

Guidance and manufacturer’s declaration - electromagnetic emissions

The unit is intended for use in the electromagnetic environment specified below. The customer or the user of the unit should assure that it is used
in such an environment.

Emissions test

Compliance

Electromagnetic Environment guidance

RF emissions

CISPR 11

Group 1

The unit uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.

RF emissions

CISPR 11

Class B

The unit is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/
flicker emissions

IEC 61000-3-3

Complies