[Read the Important Safety Rules section before operating this equipment.]

17

MN165-1 Rev G

Problem

Probable Cause

Solution


Condensation
collects in the
oxygen tubing
when you use
a humidifier bottle


Unit not properly
ventilated.
Elevated operating
temperatures.


Make sure unit is positioned away from
curtains or drapes, hot air registers or
heaters, Be certain to place the unit so all
sides are at least 1 foot (30 cm) away from
a wall or other obstruction. Do not place the
unit in a confined area.
Refill humidifier bottle with COLD water.
DO NOT OVERFILL. Allow oxygen
tubing to dry out, or replace with new
tubing.

Intermittent alarm
sounds at one
second intervals.

See section 2.6

‘Safety Features’ for

a description of

auditory indicators.

Set I/0 power switch to 0 position, use your
reserve oxygen supply (if provided), and
consult your Equipment Provider
immediately.

All other problems.

Set I/0 power switch to the 0 position, use
your reserve oxygen supply (if provided),
and consult your Equipment Provider
immediately.

Oxygen

Concentration

Indicator remains

lit (amber)

Oxygen

concentration is

82% (±3%).

Set I/0 power switch to the 0 position, use
your reserve oxygen supply (if provided),
and consult your Equipment Provider
immediately.