[Read the Important Safety Rules section before operating this equipment.]

12

MN165-1 Rev G


2.5 Materials in direct or indirect contact with the patient

o

Concentrator Cabinet (all parts) …………… NOVA PS6201 Super High Impact Polystyrene

o

Gas Outlet, Nozzle ……………………….

Aluminum, Black Anodized

o

On / Off Switch ………………………..

DuPont FR50

o

Hour Meter ………………………….........

Acrylonitrile Butadiene Styrene (ABS) Plastic &

Acrylic
o

Flowmeter ………………………………..

ABS (STAT-TECH AS1000)

o

Gas outlet ……………………………........

Brass

o

Circuit Breaker …………………………...

Button - Melamine or Urea Formaldehyde,

Mounting Bushing – Polyamide 66 (Nylon)

o

Unit Labels ……………………………….

Lexan

o

Power Cord………………………………..

Polyvinyl Chloride (PVC), Metal

o

Cord Strain relief …………………………

Nylon

o

Cord Wrap ………………………………..

Velcro

o

Dust Filter ………………………………...

Polyester

o

Humidifier Strap …………………….........

Yarn, Rubber

o

Humidifier Bottle ………………………...

Top – Acrylonitrile Butadiene Styrene (ABS)

………………………… Bottle – Polypropylene (PP), Diffuser -Polyvinyl

Chloride (PVC)

o

Cannula (Tubing)…………………………

Polyvinyl Chloride (PVC),


2.6 Safety Features:

Ø Compressor Motor: Thermal safety is ensured by a thermostat situated in

the stator winding (135

°C/ 275 °F). A safety relief valve is fitted to the

compressor outlet and is calibrated to 280 kPa (40 psig).

Ø Power Failure : In the event the unit is operating and a loss of power occurs,

an audible single - pulse intermittent alarm will activate.

Ø Oxygen Monitor: The oxygen monitor detects any drop in concentration

below 82% (±3%) and activates a visual alarm (amber LED indicator on front
panel). If the concentration continues to operate below 82%, an audible 2 -
pulse intermittent alarm will also activate.

Ø Hi Pressure: An over-pressure safety feature prevents potential damage to

device components at a pressure of 207 kPa (30 psig (± 2 psig)). An audible
4-pulse intermittent alarm will activate.

Ø Low Pressure: An audible 3- pulse intermittent alarm will activate if the

device pressure reduces to

35 kPa (5 psig (± 1 psig)).

Ø No Flow Warning : An obstruction to the flow of oxygen, such as a pinch or

kink in the delivery cannula, will activate an audible 6- pulse intermittent
alarm.

Ø

Product Filter: ≥ 10 micron