beautypg.com

Visionaire, 2 and visionaire – AirSep MN165-1 G User Manual

Page 21

background image

[Read the Important Safety Rules section before operating this equipment.]

18

MN165-1 Rev G

6.0 PRODUCT SPECIFICATIONS





VisionAire

2 and VisionAire

3

Flow Specification

s

(1)

½ lpm – 3 lpm

⅛ lpm – 2 lpm

±10% of indicated setting, or 200 mL, whichever is greater

(1)

Oxygen Concentration

(1)

90% (+ 5.5% / - 3%)

Electrical Requirements

115 VAC / 60 Hz, 3.0 A
230 VAC / 50 Hz, 1.5 A

230 VAC/ 60 Hz, 1.5 A

Power Consumption

VisionAire 2 = 200 Watts (+/- 10%)
VisionAire 3 = 175 Watts (+/- 10%)

Sound

40 dba

Dimensions

14.1 in. W x 11.5 in. D x 20.8 in. H
(35.8 cm W x 29.2 cm D x 52.8 cm H)

Weight

30 lbs (13.6 kg)

Electrical Shock

Protection

Class II, Type B

Environmental Limit

Conditions

(2)

Operating conditions: 5

o

C to 40

o

C (41

o

F to 105

o

F ) at

altitudes up to 10,000 ft.(523mmHg) above sea level.
Storage: -20

o

C - 60

o

C (-4

°F - 140°F)

Relative Humidity: up to 95% RH non - condensing

(1) Performance based at 21

°C (70°F) at nominal operating pressure range: 0-7 kPa

(0-1 psig backpressure).

Maximum outlet pressure : 55 kPa. ( 7.98 psig)

(2) Operating outside of these operational specifications can limit the concentrators ability