AirSep MN104-1 D User Manual
Page 32
28
Solution
Replace battery or connect to a DC or an
AC outlet within 5 minutes.
Replace battery or connect to a DC or an
AC outlet immediately.
Reduce activity, and consider locating
another source of oxygen.
Reduce activity, and consider locating
another source of oxygen.
Change to another source of oxygen, and
contact your Equipment Provider.
Change to another source of oxygen, and
contact your Equipment Provider.
Change to another source of oxygen, and
contact your Equipment Provider.
Probable Cause
Battery requires charging.
Battery voltage is too low to operate the
LifeStyle unit.
Breathing rate is approaching the
threshold of the LifeStyle unit's capacity.
A general malfunction could be imminent.
Breathing rate has exceeded the capacity
of the LifeStyle unit.
A general malfunction has occurred.
Problem
Intermittent alarm condition
(1-second warning sounds every 5
seconds), and the (yellow) BATTERY light
illuminates intermittently.
Rapid alarm condition
(½-second alarm sounds every ½
second), and the (yellow) BATTERY light
illuminates intermittently.
Rapid alarm/pause/alarm condition
(three ½-second warnings sound every 5
seconds), and the (red) ALARM light
illuminates intermittently.
Rapid alarm condition
(½-second alarm sounds every ½
second), and the (red) ALARM light
illuminates intermittently.
Rapid alarm condition
(½-second alarm sounds every ½
second), and the (red) ALARM illuminates
continuously.
All other problems.