beautypg.com

Cleaning, care, and proper maintenance, Compressor – AirSep MN104-1 D User Manual

Page 26

background image

22

Cleaning, Care, and Proper Maintenance

Compressor

The compressor must be replaced after approximately 3,000 hours of use. When you remove the air intake/gross
particle filter for cleaning, take note of the reading on the hour meter positioned beneath this filter. (See Figure 8.) When
the reading approaches 3,000 hours, contact your Equipment Provider for replacement of the compressor. At this
time, the Equipment Provider should also replace the product filter and felt filter.

Figure 8: Hour Meter Position

Hour Meter