beautypg.com

AirSep MN104-1 D User Manual

Page 25

background image

Audible Alarm

Continuous:

Beep

Intermittent:

Beep, beep, beep...

Rapid intermittent:

Beep, beep, beep...

Rapid intermittent:

Beep, beep, beep...

Light

(Red) ALARM;

continuous light

(Yellow) BATTERY;

intermittent light

(Red) ALARM;

intermittent light

(Red) ALARM;

continuous light

Indicates

No breath detected by the unit

for a predetermined time

period.

Battery voltage is too low to

operate LifeStyle.

Breathing rate is exceeding

the capacity of the LifeStyle

unit.

General malfunction of the

LifeStyle unit has occurred.

Action

Check the cannula connection.

Ensure that you are breathing through

your nose. If the alarm persists,

contact your Equipment Provider.

Replace the battery, or connect the

LifeStyle unit into a DC auto adapter

outlet or an AC outlet immediately.

Reduce activity, and/or locate and use

another source of oxygen. Contact

your Equipment Provider.

Turn off the unit. Change to another

source of oxygen, and contact your

Equipment Provider.

21

LIFESTYLE PORTABLE OXYGEN CONCENTRATOR

Status

Alarm

Alarm

Alarm

Alarm