LIFESTYLE PORTABLE OXYGEN CONCENTRATOR
27
Problem
LifeStyle does not operate when the flow
selector is NOT in the OFF position.
A continuous alarm sounds and the (red)
ALARM light illuminates continuously.
Probable Cause
Battery is discharged.
Battery is not installed correctly.
Malfunction.
LifeStyle has not detected a breath for a
predetermined time period.
Solution
Power the unit through the DC auto
adapter/outlet, or an AC adapter/outlet, or
replace with a fully charged battery.
Remove the battery and reinsert it,
ensuring that it is in the correct position.
Contact your Equipment Provider, and
change to a reserve oxygen supply.
Check the cannula connection.
Ensure that cannula tubing is not kinked.
Make sure that the cannula is positioned
properly and that you are breathing
through your nose.
Troubleshooting
If your LifeStyle Portable Oxygen Concentrator fails to operate properly, consult your Equipment Provider, and then
refer to the troubleshooting chart on the following pages for probable causes and solutions.
Do not attempt any maintenance other than the possible solutions listed below.