Turning On the Unit for Proper Oxygen
Usage

Note: Do not use a mask with the ImPulse Elite unit. A mask
may not conform to the face properly enough for the unit to
fully sense inhalation.

Note: Do not use pediatric or low flow cannulas with the
ImPulse Elite unit as their smaller tubing diameters negatively
affect delivery volumes.

Note: Use ImPulse Elite during sleep only under the
recommendation of a qualified clinician.

1. Check the energy level of the battery each time

you use the unit.

Refer to “Testing and Replacing Battery” for

step-by-step instructions.

2. Check the oxygen cylinder setup and the

cylinder contents gauge.

If you need to change cylinders, which includes

removal and reconnection of the ImPulse Elite

unit, refer to Oxygen Cylinder Connection for

step-by-step instructions.

3. Make sure the PULSE FLOW / CONTINUOUS

FLOW switch is moved completely right (for

pulse flow setting).

4. Set the ON/OFF switch to the ON position.

All three lights illuminate for less than 1

second. The ImPulse Elite unit automatically

calibrates and reaches sensitivity within

approximately 15 seconds. Then the green light

shuts off, and the unit emits a short burst of

oxygen. This indicates ImPulse Elite is ready

for use.

5. Connect your cannula to the oxygen outlet.

(You may use up to a 25 ft [7.5 m] cannula, if

desired.) (Figure 11.)

Figure 11

6. Turn the cylinder valve slowly one full turn

counter-clockwise to begin oxygen flow.

7. Make sure the cylinder contents gauge on the

ImPulse Elite unit registers within the green

shaded/green zoned areas.

This indicates an adequate oxygen supply in

the cylinder. (If the cylinder contents gauge

indicates the red REFILL zone, the cylinder

needs to be replaced.)

Listen for gas leaks coming from the ImPulse

Elite and oxygen cylinder. If you hear a leak,

turn off the cylinder valve and refer to the

Troubleshooting Guide on page 11.

8. Position the nasal cannula with the prongs

inserted into your nostrils.

9. Set the flow selector between 1 and 6 depending

on your prescribed dosage (Figure 12).

Figure 12

Turning Off the Unit

1. Turn the oxygen cylinder valve fully clockwise

to stop oxygen flow.

2. Set the ON/OFF power switch on the ImPulse

Elite unit to the OFF position.

Note: If you do not set the unit’s ON/OFF power switch to OFF
when the unit is not in use, the alarm sounds until the battery
fully discharges.