How to Operate the Unit
Oxygen Cylinder Connection
WARNING: HIGH PURITY OXYGEN
PROMOTES RAPID BURNING. DO NOT
ALLOW SMOKING OR OPEN FLAMES WITHIN
FIVE FEET OF THIS DEVICE, THE PORTABLE
OXYGEN CYLINDER, OR ANY OXYGEN-CARRYING
ACCESSORY.
WARNING: USE NO OIL, GREASE, OR OTHER PETROLEUM-
BASED PRODUCTS ON OR NEAR THE DEVICE OR ANY
OXYGEN SOURCE.
WARNING: USE THIS DEVICE AND THE OXYGEN CYLINDER
ONLY WITH CLEAN, DRY, OIL-FREE HANDS.
WARNING: OXYGEN IN THE CYLINDER IS UNDER HIGH
PRESSURE. DO NOT OPEN THE CYLINDER VALVE WITHOUT
THE IMPULSE ELITE UNIT PROPERLY CONNECTED TO THE
CYLINDER.
Your Equipment Provider will review proper cylinder
care and connection procedures with you the first time
you operate your unit.
Refer to the procedure on the right should you need
assistance with connecting your ImPulse Elite unit to
an oxygen cylinder.
To connect the ImPulse Elite to any post-valve
cylinder:
1. Loosen the ImPulse Elite securing knob. To
ensure a leak-tight seal between the cylinder
valve and the unit, make sure the sealing
washer is present in the center of the unit before
installation (Figure 3).
Figure 3
2. Align the two parallel alignment pins
(Figure 4) on the unit with the holes on the neck
of the cylinder (Figure 5).
Figure 4
Figure 5
3. Push the alignment pins into the holes on the
neck of the cylinder, and hold the unit in place
against the cylinder with one hand while you
turn the unit’s securing knob clockwise with
your other hand until the unit is secured firmly
on the cylinder (Figure 6).
Figure 6
Sealing Washer
Alignment Holes
Cylinder
Valve Toggle
Alignment
Pins