beautypg.com

Testing and replacing battery – AirSep MN092-1 D User Manual

Page 7

background image

ImPulse Elite User Manual — 7

Sealing Washer

Testing and Replacing Battery

ImPulse Elite requires just one 1.5 volt “D” size

alkaline battery (Energizer recommended) for

full operation. Battery life is dependent upon the

predetermined unit settings, your breathing rate, and

the length of daily ambulation time. Most patients

experience 10–18 months of battery life. AirSep

strongly recommends that you maintain a spare

battery in your carrying case for replacement in this

unit as necessary.

To install or replace the battery, follow the steps

below:

1. Make sure the ON/OFF power switch is set

to the OFF position. (It is not necessary to

disconnect the cannula or remove the ImPulse

Elite from the cylinder, although you may find

it easier to replace the battery within the battery

compartment door if the unit is first removed

from the cylinder.)

2. If the unit is not connected to the cylinder,

simply hold the unit with one hand while you

press in and lift up on the battery compartment

latch with your other hand. (Figure 7).

Figure 7

If the unit is connected to the cylinder, steady

the unit by grasping the neck of the cylinder

just below the unit with one hand while you

press in and lift up on the battery compartment

latch with your other hand (Figure 8).

Figure 8

3. Insert a fully charged battery into the inside

door of the battery compartment. Make sure

the battery is properly positioned to match the

correct positive and negative polarity markings

(Figure 9).

Figure 9

“+”Positive
Polarity

“-” Negative
Polarity

To test the battery to ensure that there is proper

power available to operate the unit:

1. Make sure there is a battery in the unit’s battery

compartment and it is properly positioned to

match the correct positive and negative polarity

markings.

2. Make sure the ON/OFF power switch is set to

the ON position.

3. Press and hold down firmly the center of the

BAT. TEST button with your finger. (Figure

10.) (You can test the battery regardless of

whether or not a cannula is connected to the

oxygen outlet.)

Figure 10

The green pulse flow indicator illuminates to

indicate a good battery. If the battery requires

replacement, the green PULSE FLOW and

yellow LOW BATTERY indicators illuminate.