7.0 Specifications
Performance Data:
Accuracy
1
± 0.5% of Reading
Optional ± 0.25% of Reading
Resolution:
Infinite. lmited only by output noise level
(
≤0.005% FS)
Thermal effect:s
Compensated Range:
0 to +50°C
Zero Shift:
< ± 0.25% FS/50 °C
Span Shift:
< ± 1.35% Rdg/50°C
Long Term Stability
±0.5% FS/Yr, excluding environmental effects
Proof Pressure:
45 psia
Environmental Data:
Operating Temperature:
0 to +80°C
Storage Temperature:
-20 to +85°C
Electrical Data (Voltage):
Connectors:
9 Pin D-Sub, 15 Pin D-sub on 6” Pigtail, or
5 pinTerminal Strip
Excitation/Output:
12 to 30 VDC for 0 to 10 VDC output
9 to 30 VDC for 0 to 5 VDC output
Output Load:
>10 K
Ω load
Power Consumption:
< 200 mW
Time Constant:
< 20 ms
EMC Performance:
Complies with EMC Directive 2004/108EC
Physical Description:
Case:
Stainless steel
Vacuum Fittings:
0.25” or 0.5” OD Tube; other fittings available;
see specification sheet
Wetted Materials
2
:
Inconel®
Measurement cavity volume
3
: <6.0 cc
Weight (approx):
<250 g
Agency Compliance
EU RoHS Compliant
CE
Notes:
1. Accuracy is expressed as % of reading. However, near Zero, the accuracy is limited by the resolu-
tion of the instrument. So, the accuracy is more correctly stated as the greater of ±0.5% reading
or ±0.01% of Full Scale.. (For the optional accuracy, this becomes the greater of ±0.25% reading or
±0.01% FS).
2. Wetted material is for 0.5” tube option only. Other flange options will add stainless steel.
3. Maximum cavity volume including the 0.5” OD tube volume of 0.26 in
3