6.0 Maintenance & Troubleshooting
There are no general maintenance requirements for the Model 730 other than
periodic zero adjustment. If the unit fails to operate when received or if the
unit appears damaged, notify Setra Systems or your supplier immediately.
Retain packaging materials for inspection. Do not use if damaged. If the Model
730 is not going to be used immediately then replace the protective flange
cover and store in suitable conditions described in Section 2.
If no obvious damage has occurred, a few simple checks can be made to verify
proper installation. Table 2 shows the solution to common problems with the
installation. If none of these problems/solutions are applicable, then please
contact a Setra Systems applications engineer for further assistance.
Table 2: Common Installation Problems & Solutions
Problem
Cause
Solution
Incorrect or no supply voltage
Ensure power supply is
No signal output
used as specified in Section 3.
Readout display short circuit or
Ensure impedance of readout
incorrect impedance
unit is > 10 k
Ω
Incorrect wiring
Ensure wiring conforms to
Signal output
diagrams in Section 3.0
reads
Potential difference between
Ensure common chassis
over-range
chassis ground of unit , power
ground between unit, power
supply and readout / display
supply and display.
Incorrect zero adjustment
Adjust zero per section 5.1
Signal output
Readout display incorrect
Ensure impedance of
reads
impedance
readout unit is > 10 k
Ω .
under-range
Incorrect wiring polarity to
Ensure wiring conforms to
readout display
diagrams in Section 3.0
Chassis ground not connected
Ensure common chassis
ground between unit, power
supply and display.
Unstable signal
Unstable or unregulated
Use a regulated power supply
supply
power supply
as specified in Section 3.0.
Electrical noise on chassis
Ensure common chassis
ground
ground between unit, power
supply, and display.