Factory default settings, E-m-hf3-v1_22, 2 factory default settings – ROTRONIC HF3 User Manual
Page 9
E-M-HF3-V1_22
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF3 Transmitters and Thermo-
hygrostats: User Guide
Instruction Manual
Document Type
Page 9 of 27
Document title
© 2009-2011; Rotronic AG E-M-HF3-V1_22
4.2 Factory default settings
Note: Configuration of the HF3 by the user and access to its functions requires a PC with the ROTRONIC
HW4 software (version 2.1.1 or higher) installed. Service cable AC3006 or AC3009 is used to connect the
HF3 service connector to a USB port of the PC (see Maintenance > Service Cable).
Configurable Settings
Applicability
Factory default
Unit system (Metric or English)
HF32, HF33, HF34
As per ordering code
Analog signal type (4…20 mA or other)
HF33
As per ordering code
Psychrometric calculation
HF32, HF33, HF34
As per ordering code
Output 1 parameter, scale and unit
HF32, HF33
As per ordering code (%RH or DP)
Output 2 parameter, scale and unit
HF32, HF33
Temperature, unit as per ordering code
Relay 1 set-point potentiometer adjust. span
HF34
As per order (%RH or DP)
Relay 2 set-point potentiometer adjust. span
HF34
As per order (temperature)
Display resolution (optional)
HF32-S, HF33-S
1 decimal
Display backlight (optional)
HF33-S
Always on
Displayed parameters
HF32, HF33
As per ordering code
Trend indicator (display)
HF32-S, HF33-S
Enabled
Device name
HF32, HF33, HF34
Instrument model
Functions
Applicability
Factory default
Humidity / temperature adjustment
HF32, HF33, HF34
Device write protection
HF32, HF33, HF34
Disabled
Limit humidity output to 100 %RH
HF32, HF33
Enabled
Out-of-limit value digital / display alarm
HF32-S, HF33-S
Disabled
Data recording
HF32, HF33, HF34
Enabled (loop mode
– 10 min. interval)
Automatic humidity sensor test
HF32, HF33, HF34
Disabled
Humidity sensor drift compensation
HF32, HF33, HF34
Disabled
Fail safe mode
HF32, HF33, HF34
Disabled
Simulator mode
HF32, HF33
Disabled
o
For a detailed description of all AirChip 3000 / HF3 main functions see document E-T-AC3000-DF-V1
o
Instructions regarding the configuration of the HF3 and access to its functions are provided in the
following manuals:
E-M-HW4v3-Main
E-M-HW4v3-F2-002
E-M-HW4v3-DR-001
E-M-HW4v3-A2-001
E-M-AC3000-CP
.
o
The factory default setting for dew / frost point calculation is frost point below freezing