7
Operation
7.1 Minimum load requirements for the HF33 with voltage outputs
The following requirements apply to any external device or circuit connected to the HF33 transmitter with
voltage outputs:
In the situation where the external device uses an internal pull-up resistor the value of this resistor should
meet the requirements shown below. It is also necessary to add a pull-down resistor RL connected to ground
in order to be able to read 100% of the range of the HF33 voltage output.
HF33 output signal
Input resistance RL
0…1V
>=1kOhm
0…5V
>=5kOhm
0…10V
>=10kOhm
HF33 signal
VCC
R pull-up
RL
0…1V
3.3V
≥ 250 kOhm
1 kOhm
HF33 signal
VCC
R pull-up
RL
0…1V
5.0V
≥ 400 kOhm
1 kOhm
0…5V
5.0V
≥ 400 kOhm
5 kOhm
HF33 signal
VCC
R pull-up
RL
0…1V
10.0V
≥ 1 MOhm
1 kOhm
0…5V
10.0V
≥ 1 MOhm
5 kOhm
0…10V
10.0V
≥ 1 MOhm
10 kOhm
RL
GND
HF33
Voltage Output
R Pull-up
RL
GND
VCC
HF33
Voltage Output