Hygrolog hl20 and hl21 data logger: user guide – ROTRONIC HygroLog HL20 User Manual
Page 15
IN-E-HL20-V1_12
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroLog HL20 and HL21 data logger:
User Guide
Instruction Manual
Document Type
Page 15 of 17
Document title
© 2010-2011; Rotronic AG
IN-E-HL20-V1_12
Temperature measurement
HL20 / HL21
Sensor
Pt100 RTD, IEC 751 1/3 class B
Measuring range
-10...60°C
Measurement accuracy at 23 °C
±0.2 °C
Repeatability
0.05°C
Long term stability
< 0.1°C / year
Response time
Typical 10 min. (depends on air flow)
Calculated parameters
HL20
Psychrometric calculations
Dew or frost point (user configurable setting)
Start-up time and data refresh rate
HL20 / HL21
Start-up time
1.9s (typical)
Data refresh rate
1.7s (typical)
Data recording and download
HL20 / HL21
Recorded parameter(s)
HL20: relative humidity and temperature, HL21: temperature
Recording media
Internal EEPROM non volatile, not removable
Recording capacity
HL20: 20,000 value pairs, HL21: 20,000 values
Real time clock
Yes, adjustable via PC
Logging modes
Start-stop (recording stops when memory is full
Loop: oldest data is dumped to make room for latest data
Log interval
Min. 5 sec, up to1 hour in 5 sec. increments
Interface for data download
Internal connector, located under protective cover
Service and data connector
HL20 / HL21
Interface type
UART (Universal Asynchronous Receiver Transmitter)
Maximum service cable length
5 m (16.4 ft)
General specifications
HL20 / HL21
Optional display
LC, 1 or 2 decimals resolution, backlight, alarm and low battery indication
LED indicator
5 sec, green flash when recording data
5 sec. red flash for out-of-limit value or logger problem
Housing material
ABS
Housing protection grade
HL20: IP 40, HL21: IP65
Overall dimensions
130 x 72 x 32 mm (5.1 x 2.8 x 1.3
”)
Weight
About 210 g (7.4 oz)
Conformity with standards
HL20 / HL21
CE / EMC immunity
EMC Directive 2004/108/EG: EN 61000-6-1: 2007, EN 61000-6-2: 2005
EN 61000-6-3: 2007, EN 61000-6-4: 2007
Solder type
Lead free (RoHS directive)
FDA / GAMP directives
compatible