Procedure for adjusting the data logger with the ROTRONIC HW4 software:

Use cable AC3006 to connect the service connector of the data logger to the USB port of

a PC with the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver
must be installed on the PC as explained in the HW4 manual E-M-HW4v3-Main

Start HW4 on the PC and search for the data logger (HW4 Main Menu Bar > Devices and

Groups > Search for USB Masters).

After finding the data logger with HW4, expand the device tree to see the data logger

functions. Select Probe Adjustment.

For further instructions see HW4 manual E-M-HW4v3-A2-001

4.2 Firmware updates

Firmware updates will be available on the ROTRONIC website for downloading. Firmware files
are given a name that shows both to which device the file applies and the version number of the
firmware. All firmware files have the extension HEX. Procedure for updating the firmware:

Use cable AC3006 to connect the service connector of the data logger to the USB port of

a PC with the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver
must be installed on the PC as explained in the HW4 manual E-M-HW4v3-Main

Copy the firmware update file from the ROTRONIC website to the PC.
Start HW4 software on the PC and search for the data logger (HW4 Main Menu Bar >

Devices and Groups > Search for USB Masters).

After finding the data logger, expand the device tree to see the data logger functions.

Select Device Manager. In the Device Manager menu bar select Tools > Firmware
Update. For instructions see document E-M-HW4v3-F2-018

5.

Technical data

5.1 Specifications

General

HL20 / HL21

Device type

Humidity-temperature data logger with internal sensor(s)

Battery type

3 x AA (1.5 V) alkaline

Battery life time

Depends on log interval and other settings. Examples based on a 30 minutes log
interval:

-

Models w/o LC display: 20 months

-

Models with LC display, display sleep function set to 1 minute: 13 months

Low battery indication

Yes (with HW4 software or with optional display)


Humidity measurement

HL20

Sensor

ROTRONIC Hygromer

®

IN1

Measuring range

0…100 %RH

Measurement accuracy at 23 °C

±0.8 %RH

Repeatability

0.3 %RH

Long term stability

< 1 %RH / year

Response time

Typical 3 min. (depends on air flow)