Bandwidth, Current setting – FUTEK CSG110 (JM-2A / JM-2AD)-Series User Manual
Page 8
- 8 -
Tel: (949) 465-0900
Fax: (949) 465-0905
Toll Free: (800) 23 FUTEK
1 UP
0.5
5 UP
2.5
DOWN N / A
DOWN
N / A
2 UP
1
6 UP
3
DOWN N / A
DOWN
N / A
3 UP
1.5
7 UP
4
DOWN N / A
DOWN
N / A
4 UP
2
8 UP
10
DOWN N / A
DOWN
N / A
SW3: Sensitivity (mV/V)
Bandwidth
There are two bandwidth settings available on the CSG110: 1 kHz (default), and 10 kHz.
To select the bandwidth, simply flip the DIP switch to the appropriate configuration.
Ex. If your application requires a higher bandwidth in order to account for the higher
frequency signals, such as those from dynamic applications, then you would want to
use the higher setting bandwidth.
UP
10
(3)
DOWN
1
SW4: Bandw idth
Current Setting
(4)
There are four current output settings available on the CSG110: 0-16 mA, 0-20 mA, 4-20
mA (default), and 5-25 mA. To select the current, simply flip the DIP switches to the
appropriate configuration.
SW5
SW6
SW7
Input Range (V) Output Range (mA)
DOWN
DOWN
DOWN
0-10
4-20
DOWN
DOWN
UP
0-10
5-25
UP
DOWN
DOWN
0-10
0-16
UP
DOWN
UP
0-10
0-20
DOWN
UP
DOWN
0-5
4-20
DOWN
UP
UP
0-5
5-25
UP
UP
DOWN
0-5
0-16
UP
UP
UP
0-5
0-20
Note: Only available with current output option.