beautypg.com

Dwyer 3200G User Manual

Page 17

background image

6.3 Hardware Diagnostics
If there is a failure dispite a diagnostic message on the HHT, Table
6.1 can help troubleshoot the problem.

Page 16

Symptom
Transmitter does not
Communicate with HART

®

Communicator

High Output

Erratic Output

Low Output or No Output

Potential Source
Loop Wiring

Sensor Input Failure

Loop Wiring

Power Supply

Electronics Module

Loop Wiring

Electronics Module

Sensor Element

Loop Wiring

Electronics Module

Corrective Action
• Check for a 250-550 ohms resistance

between the power supply and HHT.

• Check for adequate voltage to the transmitter

(the transmitter requires 11.9 ~ 45 Vdc).

• Check for intermittent shorts, open circuits, and

multiple grounds.

• Connect HHT and enter the transmitter test

mode to isolate a sensor failure.

• Check for dirty or defective terminals,

interconnecting pins, or receptacles.

• Check the output voltage of the power supply

at the transmitter terminals. It should be 11.9
to 45 Vdc.

• Connect HHT and enter the transmitter test

mode to isolate module failure. Check the
sensor limits to ensure the calibration
adjustments are within the sensor range.

• Check the output voltage of the power supply

at the transmitter terminals. It should be 11.9
to 45 Vdc.

• Check for intermittent shorts, open circuits, and

multiple grounds.

• Check for proper polarity at the signal terminals.
• Connect HHT and enter the transmitter test

mode to isolate an electronics mode failure.

• Connect HHT and enter the transmitter test

mode to isolate a sensor failure.

• Is the PV out of range.
• Check for adequate voltage to the transmitter

(the transmitter requires 11.9 ~ 45 Vdc).

• Check for intermittent shorts, open circuits, and

multiple grounds.

• Check polarity of signal terminal
• Check the loop impedance.
• Connect HHT and check the sensor limits to

ensure calibration adjustments are within the
sensor range.

Table 6.1 Troubleshooting