Instructions for Removing the Local Register
1 To remove the register cover and clamp ring, remove the seal screw that acts as a hinge pin Remove the clamp ring by
spreading it slightly apart at the open side to provide clearance so it can be lifted off the meter head
2 Slide out the register lock plate, located between the register base and the meter head at the entrance side of the
rotor cage
3 Rotate the register 45 degrees in either direction to release it from the bayonet lock detents on the head assembly
4 Release the retaining ring surrounding the register extension tube inside the rotor cage Using a Truarc right-angle
pliers, expand the retaining ring and, at the same time, pull the register tube out of the head unit, removing the
register assembly
Figure 8: Register and Operating Head Assembly
(2" and 3" Meters shown)