beautypg.com

Pc-extsen-d pin-outs – Alicat P Series Pressure Gauge User Manual

Page 82

background image

82

PC-EXTSEN-D Pin-Outs

The following pin-out diagram is applicable to all standard PC-EXTSEN-D

devices. If your device was ordered with custom pin-out requirements, please

contact Alicat for assistance.

PC-EXTSEN-D Standard 8 Pin Mini-DIN Pin-Out

Pin Functi on

Mini-DIN

cable color

1

Inacti ve or 4-20mA Primary Output Signal

Black

2

External Sensor Signal Input*

Brown

3

RS-232 Input Signal

Red

4

Analog Input Signal

Orange

5

RS-232 Output Signal

Yellow

6

0-5 Vdc (or 0-10 Vdc) Output Signal

Green

7

Power In

Blue

8

Ground (common for power, communicati ons and signals)

Purple

*If you are using the PC-EXTSEN-D-ISC’s 6 pin locking connecti on to receive the

external sensor input signal do not wire any signal into pin 2 of the Mini-DIN.

Pin 7 which is normally uti lized to provide power into the Alicat device can

be used to provide power to your external sensor. Please make sure that

the power you are providing to the device is compati ble with the power

your sensor can accept before choosing to wire power in this manner.

CAUTION! Do not connect power to pins 1 through 6 as permanent

damage can occur!

POWER AND SIGNAL CONNECTIONS

Power can be supplied to your controller through either the power jack or the 8 pin

Mini-DIN connector.

An AC to DC adapter which converts line AC power to DC voltage and current as

specifi ed below is required to use the power jack.

A 2.1mm, positi ve center, 12-30 Vdc AC/DC adapter rated for at least 250 mA is

required to use the adapter jack in a PC-Series controller.

A 2.1mm, positi ve center, 24-30 Vdc AC/DC adapter rated for at least 500 mA is

required to use the adapter jack in a PCR-Series controller.

NOTE: 4-20mA analog output requires at least 15 Vdc.