Technical Data for PCS and PCRS Single Valve Pressure Controllers

Standard Specifications (Contact Alicat for available options.)

Performance

PCS Controllers

PCRS Controllers

Full scale pressure Standard Accuracy

± 0.25%

Full scale pressure High Accuracy Option

± 0.125%

Repeatability

± 0.08% Full Scale

Zero Shift and Span Shift

0.02% Full Scale / ºCelsius

Operating Range / Turndown Ratio

0.5% to 100% Full Scale / 200:1 Turndown

Excess Pressure

102.4% FS Controllable

Burst Pressure

3 X Full Scale

Typical Response Time

1

100 ms (Adjustable)

Warm-up Time

< 1 Second

1. Volumes, feed pressures, exhaust pressures and line sizing will determine the limits of response times.

Operating Conditions

PCS Controllers

PCRS Controllers

Gas Compatibility

Compatible with all non-corrosive gases and select aggressive gases

1

Operating Temperature

-10 to +50 ºCelsius

Common Mode Pressure

(Differential Pressure Units Only)

150 psig

Mounting Attitude Sensitivity

None

Valve Type

Normally Closed

Ingress Protection

IP40

Wetted Materials

303SS, 430FRSS, 316LSS, FFKM (Kalrez) standard; Viton, EPDM, Buna,

Neoprene as needed for some gases.

If your application demands a different material, please contact Alicat.

1. In addition to all non-corrosive gases, PC & PCR controllers are configured to operate with the following

aggressive gases: Ammonia, Hydrogen Sulfide, Nitric Oxide, Nitrogen Dioxide, Nitrogen Triflouride, Propylene.

The following gases are available upon request: Refrigerant gases to 100% (Refrigerant gases my require

custom seals, consult Alicat.) Other gases to 1000 ppm in an inert carrier. If your application requires another gas

or gas mixture, please contact Alicat. PCS and PCRS devices may be used with chemically compatible liquids.

Please contact Alicat for technical assistance if your application involves fluids.

Communication / Power

PCS Controllers

PCRS Controllers

Monochrome LCD or Color TFT Display

with integrated touchpad

Displays Pressure

Digital Output Signal

1

Options

RS-232 Serial / RS-485 Serial / PROFIBUS

3

Analog Output Signal

2

Options

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Optional Secondary Analog Output Signal

2

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Electrical Connection Options

8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking

Supply Voltage

12-30 Vdc (15-30 Vdc for 4-20 mA outputs)

24-30 Vdc

Supply Current

0.250 Amp

0.750 Amp

1. The

Digital Output Signal communicates Pressure

2. The

Analog Output Signal and Optional Secondary Analog Output Signal communicate Pressure

3. If selecting

PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See

PROFIBUS specifications for PROFIBUS supply voltages and currents. (

www.alicat.com/profibus)

Mechanical Specifications

Pressure Product

Mechanical Dimensions

Process Connections

1

PCS Controllers

4.1”H x 3.6”W x 1.1”D

1/8” NPT Female

PCRS Controllers

5.7”H x 2.9”W x 5.5”D

3/4” NPT Female

1. Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and

SAE connections upon request.

Select One Unit of Measure when Ordering

PSIA

inHG

Atm

PSIG

inH

2

O

Torr

mmHG

mBar

kPa

Standard Available Ranges

PCS & PCRS Controllers

-15 psig to 0 psig

1 psid

1 psig

5 psid

5 psig

15 psid

15 psig

15 psia

30 psid

30 psig

30 psia

100 psid

100 psig

100 psia

300 psig

300 psia

500 psig*

500 psia*

*Not available on PCRS 0ther ranges available. Please contact Alicat.