beautypg.com

Standard available ranges – Alicat P Series Pressure Gauge User Manual

Page 65

background image

65

Technical Data for PCDS & PCRDS Dual Valve Pressure Controllers

Mechanical Specifications

Standard Specifications (Contact Alicat for available options.)

Performance

PCDS Controllers

PCRDS Controllers

Full scale pressure Standard Accuracy

± 0.25%

Full scale pressure High Accuracy Option

± 0.125%

Repeatability

± 0.08% Full Scale

Zero Shift and Span Shift

0.02% Full Scale / ºCelsius

Operating Range / Turndown Ratio

0.5% to 100% Full Scale / 200:1 Turndown

Excess Pressure

102.4% FS Controllable

Burst Pressure

3 X Full Scale

Typical Response Time

1

100 ms (Adjustable)

Warm-up Time

< 1 Second

1. Volumes, feed pressures, exhaust pressures and line sizing will determine the limits of response times.

Operating Conditions

PCDS Controllers

PCRDS Controllers

Gas Compatibility

Compatible with all non-corrosive gases

1

Operating Temperature

-10 to +50 ºCelsius

Common Mode Pressure

(Differential Pressure Units Only)

150 psig

Mounting Attitude Sensitivity

None

Mount with valve cylinder vertical & upright

Valve Type

Normally Closed

Ingress Protection

IP40

Wetted Materials

316LSS, 303SS, 430FRSS, FFKM (Kalrez) standard; Viton, EPDM, Buna,

Neoprene as needed for some gases. If your application demands a different

material, please contact Alicat.

1. In addition to all non-corrosive gases, PC & PCR controllers are configured to operate with the following

aggressive gases: Ammonia, Hydrogen Sulfide, Nitric Oxide, Nitrogen Dioxide, Nitrogen Triflouride, Propylene.

The following gases are available upon request: Refrigerant gases to 100% (Refrigerant gases my require

custom seals, consult Alicat.) Other gases to 1000 ppm in an inert carrier. If your application requires another gas

or gas mixture, please contact Alicat. PCS and PCRS devices may be used with chemically compatible liquids.

Please contact Alicat for technical assistance if your application involves fluids.

Communication / Power

PCDS Controllers

PCRDS Controllers

Monochrome LCD or Color TFT Display

with integrated touchpad

Displays Pressure

Digital Output Signal

1

Options

RS-232 Serial / RS-485 Serial / PROFIBUS

3

Analog Output Signal

2

Options

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Optional Secondary Analog Output Signal

2

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Electrical Connection Options

8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking

Supply Voltage

12-30 Vdc (15-30 Vdc for 4-20 mA outputs)

24-30 Vdc

Supply Current

0.250 Amp

0.750 Amp

1. The

Digital Output Signal communicates Pressure

2. The

Analog Output Signal and Optional Secondary Analog Output Signal communicate Pressure

3. If selecting

PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See

PROFIBUS specifications for PROFIBUS supply voltages and currents. (

www.alicat.com/profibus)

Dual Valve Pressure Controllers

Mechanical Dimensions

Process Connections

1

PCDS All Standard Ranges

4.1”H x 4.8”W x 1.1”D

1/8” NPT Female

PCRD All Standard Ranges

5.5”H x 10.6”W x 2.3”D

3/4” NPT Female

1. Compatible with Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and

SAE connections upon request.

Select One Unit of Measure when Ordering

PSIA

inHG

Atm

PSIG

inH

2

O

Torr

mmHG

mBar

kPa

Standard Available Ranges

PCDS and PCRDS Controllers

-15 psig to 0 psig

1 psid

1 psig

5 psid

5 psig

15 psid

15 psig

15 psia

30 psid

30 psig

30 psia

100 psid

100 psig

100 psia

150 psid

300 psig

300 psia

500 psig*

500 psia*

*Not available on PCRD. Other ranges available. Please contact Alicat.