beautypg.com

Technical data for p-series pressure gauges, Mechanical specifications, Standard available ranges – Alicat P Series Pressure Gauge User Manual

Page 51

background image

51

Technical Data for P-Series Pressure Gauges

Standard Specifications (Contact Alicat for available options.)

Performance

P-Series Gauges

Full scale pressure < 2” H2O Accuracy

Consult Factory

Full scale pressure ≥ 2” H2O

Standard Accuracy

± 0.25%

Full scale pressure ≥ 2” H2O

High Accuracy Option

± 0.125%

Repeatability

± 0.08% Full Scale

Zero Shift and Span Shift

0.02% Full Scale / ºCelsius

Operating Range / Turndown Ratio

0.5% to 100% Full Scale / 200:1 Turndown

Excess Pressure

128% FS Measurable

Burst Pressure

3 X Full Scale

Typical Response Time

1

5 ms (Adjustable)

Warm-up Time

< 1 Second

1. Volumes, feed pressures, exhaust pressures and line sizing will determine the limits of response times.

Operating Conditions

P-Series Gauges

Gas Compatibility

Compatible with all non-corrosive gases

1

Operating Temperature

-10 to +50 ºCelsius

Common Mode Pressure

(Differential Pressure Units Only)

200 psig

Mounting Attitude Sensitivity

None

Ingress Protection

IP40

Wetted Materials

302 & 303 Stainless Steel, Viton®. Silicone RTV, Silicon, Glass.

If your application demands a different material, please contact Alicat.

1. For aggressive gases, please see our PS-Series pressure Gauges. For use with water or other liquids please

contact Alicat

Communication / Power

P-Series Gauges

Monochrome LCD or Color TFT Display

with integrated touchpad

Displays Pressure

Digital Output Signal

1

Options

RS-232 Serial / RS-485 Serial / PROFIBUS

3

Analog Output Signa

l

2

Options

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Optional Secondary Analog Output Signal

2

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Electrical Connection Options

8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking

Supply Voltage

7-30 Vdc (15-30 Vdc for 4-20 mA outputs)

Supply Current

0.040 Amp

1. The

Digital Output Signal communicates Pressure

2. The

Analog Output Signal and Optional Secondary Analog Output Signal communicate Pressure

3. If selecting

PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See PROFIBUS

specifications for PROFIBUS supply voltages and currents. (

www.alicat.com/profibus)

Mechanical Specifications

Pressure Product

Mechanical Dimensions

Process Connections

1

P-Series Gauges

4.1”H x 2.4”W x 1.1”D

1/8” NPT Female

1. Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and

SAE connections upon request.

Select One Unit of Measure when Ordering

PSIA

inHG

Atm

PSIG

inH

2

O

Torr

mmHG

mBar

kPa

Standard Available Ranges

P-Series Gauges

-15 psig to 0 psig

2 inH

2

OD

2 inH

2

OG

4 inH

2

OD

4 inH

2

OG

1 psid

1 psig

5 psid

5 psig

15 psid

15 psig

15 psia

30 psid

30 psig

30 psia

100 psid

100 psig

100 psia

150 psid

300 psig

300 psia

500 psig

500 psia

Other ranges available. Please contact Alicat.