MISCELLANEOUS

Miscellaneous is accessed by pressing the MISC button on the Select Menu display.
NOTE: All Miscellaneous changes are recorded when you exit Miscellaneous.
ZERO BAND
refers to Display Zero Deadband. Zero deadband is a value below

which the display jumps to zero. This deadband is often desired to prevent electrical

noise from showing up on the display as minor flows or pressures that do not exist.

Display Zero Deadband does not affect the analog or digital signal outputs.

ZERO BAND can be adjusted between 0 and

3.2% of the sensor’s Full Scale (FS).
Press ZERO BAND. Then use SELECT to

choose the decimal with the arrow and the

UP and DOWN buttons to change the value.

Press SET to record your value. Press CLEAR

to return to zero.
Pressure Averaging may be useful to make

it easier to read and interpret rapidly

fluctuating pressures and flows. Pressure

and flow averaging can be adjusted between

1 (no averaging) and 256 (maximum

averaging). These are geometric running

averages where the number between 1 and

256 can be considered roughly equivalent to

the response time constant in milliseconds.

This can be effective at “smoothing” high

frequency process oscillations such as those

caused by diaphragm pumps.
Press PRESS AVG. Then use SELECT to

choose the decimal with the arrow and the

UP and DOWN buttons to change the value.

Press SET to record your value. Press CLEAR

to return to zero.
Setting a higher number will equal a

smoother display.

Flow Averaging: While FLOW AVG appears

on the display, this adjustment has no

application in pressure devices. Changing

it will have no effect on the operation of a

pressure device.
LCD CONTRAST: The display contrast can be

adjusted between 0 and 30, with zero being

the lightest and 30 being the darkest.
Use the UP and DOWN buttons to adjust the

contrast. Press SET when you are satisfied.
Press CANCEL to return to the MISC display.