Technical data for ps-series pressure gauges, Mechanical specifications, Standard available ranges – Alicat P Series Pressure Gauge User Manual
Page 61
61
Technical Data for PS-Series Pressure Gauges
Standard Specifications (Contact Alicat for available options.)
Performance
PS-Series Gauges
Full scale pressure Standard Accuracy
± 0.25%
Full scale pressure High Accuracy Option
± 0.125%
Repeatability
± 0.08% Full Scale
Zero Shift and Span Shift
0.02% Full Scale / ºCelsius
Operating Range / Turndown Ratio
0.5% to 100% Full Scale / 200:1 Turndown
Excess Pressure
128% FS Measurable
Burst Pressure
3 X Full Scale
Typical Response Time
1
5 ms (Adjustable)
Warm-up Time
< 1 Second
1. Volumes, feed pressures, exhaust pressures and line sizing will determine the limits of response times.
Operating Conditions
PS-Series Gauges
Gas Compatibility
Compatible with all non-corrosive gases and select aggressive gases
1
Operating Temperature
-10 to +50 ºCelsius
Common Mode Pressure
(Differential Pressure Units Only)
200 psig
Mounting Attitude Sensitivity
None
Ingress Protection
IP40
Wetted Materials
316LSS, FFKM (Kalrez) standard; Viton, EPDM, Buna, Neoprene as
needed for some gases.
If your application demands a different material, please contact Alicat.
1. In addition to all non-corrosive gases, PS Gauges are configured to operate with the following aggressive
gases: Ammonia, Chlorine, Hydrogen Sulfide, Nitric Oxide, Nitrogen Dioxide, Nitrogen Triflouride, Propylene,
Sulfur Dioxide. The following gases are available upon request: Refrigerant gases to 100% (Refrigerant gases
my require custom seals, consult Alicat.) If your application requires another gas or gas mixture, please contact
Alicat. For use with water or other liquids please contact Alicat.
Communication / Power
PS-Series Gauges
Monochrome LCD or Color TFT Display
with integrated touchpad
Displays Pressure
Digital Output Signal
1
Options
RS-232 Serial / RS-485 Serial / PROFIBUS
3
Analog Output Signa
l
2
Options
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Optional Secondary Analog Output Signal
2
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Electrical Connection Options
8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking
Supply Voltage
7-30 Vdc (15-30 Vdc for 4-20 mA outputs)
Supply Current
0.040 Amp
1. The
Digital Output Signal communicates Pressure
2. The
Analog Output Signal and Optional Secondary Analog Output Signal communicate Pressure
3. If selecting
PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See PROFIBUS
specifications for PROFIBUS supply voltages and currents. (
www.alicat.com/profibus)
Mechanical Specifications
Pressure Product
Mechanical Dimensions
Process Connections
1
PS Gauges
4.1”H x 2.4”W x 1.1”D
1/8” NPT Female
1. Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and
SAE connections upon request.
Select One Unit of Measure when Ordering
PSIA
inHG
Atm
PSIG
inH
2
O
Torr
mmHG
mBar
kPa
Standard Available Ranges
PS-Series Gauges
-15 psig to 0 psig
1 psid
1 psig
5 psid
5 psig
15 psid
15 psig
15 psia
30 psid
30 psig
30 psia
100 psid
100 psig
100 psia
150 psid
300 psig
300 psia
500 psig
500 psia
Other ranges available. Please contact Alicat.