Maintenance and recalibration – Alicat P Series Pressure Gauge User Manual
Page 44

44
MAINTENANCE AND RECALIBRATION
General: P, PC, PCR and PCD-Series Pressure Gauges and Controllers require
minimal maintenance. They have no moving parts. The single most important thing
that affects the life and accuracy of these devices is the quality of the gas being
measured. The instruments are designed to measure CLEAN, DRY, NON-CORROSIVE
gases. If your application requires an aggressive or corrosive gas, please consider
Alicat’s PS, PCS, PCRS and PCDS Series instruments (see page 60).
Recalibration: The recommended period for recalibration is once every year. A
label located on the back of the controller lists the most recent calibration date.
The controller should be returned to the factory for recalibration within one
year from the listed date. Before calling to schedule a recalibration, please note
the serial number on the back of the meter. The Serial Number, Model Number,
and Date of Manufacture are also available on the Model Info display (page 20).
Cleaning: P, PC, PCR and PCD-Series Pressure Gauges and Controllers require
no periodic cleaning. If necessary, the outside of the controller can be cleaned
with a soft dry cloth. Avoid excess moisture or solvents.
For repair, recalibration or recycling of this product, contact:
Alicat Scientific, Inc.
7641 N Business Park Drive
Tucson, Arizona 85743
USA
Ph. 520-290-6060
Fax 520-290-0109
email: [email protected]
Web site: www.alicat.com
Pressure Conversion Table
psi
1.00 =
51.7150 mmHg
psi
1.00 =
2.0360 inHg
psi
1.00 =
27.7080 inH20
psi
1.00 =
68.9480 mbar
psi 100.00 =
6.8046 atm
psi
1.00 =
51.7150 torr
psi
1.00 =
6.8948 kPa
mmHg 100.00 =
3.9370 inHg
mmHg
100 =
1.9337 psi
inHg
100 =
49.1159 psi
inH2O
100 =
3.6091 psi
mbar
100 =
1.4504 psi
atm
1 =
14.6959 psi
torr
100 =
1.9337 psi
kPa
100 =
14.5037 psi
inHg
1 =
25.4000 mmHg