Db15k – pin-out “mks” style, Db15 pin-outs – Alicat P Series Pressure Gauge User Manual
Page 78
78
DB15K – Pin-Out “MKS” Style
The following pin-out chart describes the safest and generally compatible arrangement when
connecting a non-Alicat DB15 wire to a
DB15K equipped Alicat. Not all features may be
available between brands, but the common denominators are featured in our DB15 offerings,
along with some options for customization.
DB15 Pin-Outs
If your instrument was ordered with a DB15 connection,
be sure to check the Calibration Label on the device and
reference the appropriate pin-out diagram.
Pin Number
Function
1
N/C
2
Primary Analog Signal Output
3
N/C
4
N/C
5
Power Supply Common
6
N/C
7
Power Supply (+Vdc)
8
Analog Tare (meters — when grounded)
Analog Set-Point Input (controllers)
9
Secondary Analog Signal Output / fi xed 5.12Vdc
*
10
N/C
11
Ground
12
Ground
13
RS-232 RX (receive) or RS-485 –
*
14
RS-232 TX (send) or RS-485 +
*
15
Ground
Check your device’s calibration certifi cate and user manual for the actual electrical input/
output requirements, as all instruments are custom confi gured to some extent.
NOTE: Pins 5, 11, 12 and 15 are connected together inside of the device and are common
grounding points.
N/C = Not Connected/Open (can be used for custom pin assignments – please consult factory).
* Added to allow for full use of features on Alicat devices, may not be present on host wiring.
Male Connector Front View
Female Connector Front View
8
13
14
9
2
5
8
13
14
9
2
5
7
7