12

Technical data

INPUTS

UNIVERSAL INPUT INP1

Type: single-ended, except thermocouples

Resolution:

> 14 bits

Decimal point:

0 to 3 decimals

Digital input filter:

adjustable 0,0...999,9 s

Scanning cycle:

100 ms

Linearization:

15 segments, adaptable with
BlueControl

®

Measurement value correction: 2-point or offset
Limiting frequency:

1,7 Hz

Thermocouples (Table 1)

Input resistance:

? 1 M[

Influence of source resistance: 1 úV/[

Input circuit mo

nitor

break, reversed polarity

Cold-junction compensation

w

internal,

- additional error:

typ.:

max

.:

ß_ 0,5 K
ß -2,5 K

external,
-constant reference

0...100 °C

Break monitoring

Sensor current:

1 úA

Operating sense configurable

Resistive sensors (Table 2)

Connection technique:

3-Leiter,
4-wire (not with INP2)

Lead resistance (for max. span):

max. 30 [

Input circuit monitoring:

break and short circuit

Measurement span

The BlueControl

®

software enables the internal characteristic

cu

rve for the KTY 11-6 temperature sensor to be adapted.

Divided into ranges
Physical measurement range:

0...4500 [

Current and voltage measurement (Table 3)

Span start and span:

anywhere within the
measurement range

Scaling:

freely selectable –1999...9999

Input circuit monitoring (current): 12,5% below span start (2 mA)

O

2

measurement (option)

Electromotive force measurement by means of INP1
(high-impedance mV inputs)
suitable for probes with

w

constant sensor temperature (heated probes), setting by means
of parameter

measured sensor temperature (non-heated probes),
measurement by means of INP2

ADDITIONAL INPUT INP2 (UNIVERSAL,
OPTIONAL)

Resolution:

> 14 bits

Digital input filter:

adjustable 0,0...999,9 s

Scanning cycle:

100 ms

Linearization:

as for INP1

Measurement value correction:

2-point or offset

Type:

single-ended except
thermocouples

Remaining technical data as for INP1 except:
• Voltage input ranges –10...0...+10V and -5...+5V not possible.
• Millivolt input ranges: high-impedance input for low-impedance

sources.

Technical data

TB 45

47

Thermocouple type

Measurement range

Accuracy

Typical resol.

L

Fe-CuNi (DIN)

-100...900°C

-148...1652°F

ß 2K

0,1 K

J

Fe-CuNi

-100...1200°C

-148...2192°F

ß 2K

0,1 K

K

NiCr-Ni

-100...1350°C

-148...2462°F

ß 2K

0,2 K

N

Nicrosil/Nisil

-100...1300°C

-148...2372°F

ß 2K

0,2 K

S

PtRh-Pt 10%

0...1760°C

32...3200°F

ß 2K

0,2 K

R

PtRh-Pt 13%

0...1760°C

32...3200°F

ß 2K

0,2 K

T**

Cu-CuNi

-200...400°C

-328...752°F

ß 2K

0,05 K

C

W5%Re-W26%Re

0...2315°C

32...4199°F

ß 3K

0,4 K

D

W3%Re-W25%Re

0...2315°C

32...4199°F

ß 3K

0,4 K

E

NiCr-CuNi

-100...1000°C

-148...1832°F

ß 2K

0,1 K

B*

PtRh-Pt6%

0(400)...1820°C

32(752)...3308°F

ß 3K

0,4 K

Special

-25 … +75 mV

ß 0,1%

0,01%

* Values apply from 400 °C upwards

** Values apply from –80 °C

Table 1: Thermocouple input