8.9 Set TDS Factor – P7.0
The alpha TDS 200 Controller allows the TDS factor to be selected, depending on the
application.

.

P .0

d

t s

0

0


P7.1: Viewing Temperature electrode data
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P7.0 and TDS. Press ENTER.

3.

Screen will scroll P7.1, ‘FACT.’ then TDS factor. (Default is 0.70)

4.

Press the ▲or ▼keys to change TDS factor (0.40 to 1.0). Press ENTER.

5.

Press ▲and ▼keys together (ESCAPE) to return to P7.0.



8.10 Controller Reset – P8.0
Resets the controller to factory default values.

TDS range remains unchanged based on last selected range

All calibration data and TDS factor are reset

SP 1 is reset to 10 % of full scale and LOW

SP 2 is reset to 90 % of full scale and HIGH

Hysteresis for both set points is reset to 5 % of full scale

Temperature compensation mode remains unchanged

In MTC mode, temperature value is reset to 25

o

C

Temperature Coefficient is reset to 2.1%

Normalization Temperature is reset to 25.0

o

C


NOTE: ATC temperature offset is NOT reset.