8.8 Viewing TDS / Temperature Electrode Data – P6.0
Program 6 has two “view only” options that let you check the electrode parameters for
diagnostic purposes.
P6.1: view revised cell correcting factor value of electrode
P6.2: view temperature probe offset (ATC on only)

These parameters will change each time you recalibrate the controller.

P6.0

elec

P6.1

P6.2

t.ofs

0.0

1 000

.

FCTR

Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and return
to Measurement mode.

P6.1: Viewing TDS electrode data
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P6.0 and ELEC. Press ENTER.

3.

Screen will scroll P6.1, ‘FCTR’ then cell correction factor. Each time a calibration is
performed, the value of the cell correction factor will be updated.

4.

Press ▲and ▼keys together (ESCAPE) to return to P6.0 or press ENTER to view
Temperature electrode data.


P6.2: Viewing Temperature electrode data
5.

Follow procedure from above (P6.1) until step 3 and press ENTER.

6.

Screen will scroll P6.2, ‘T.OFS’ (Temperature Offset), then amount of offset (in

o

C).

7.

Press ENTER or press ▲and ▼keys together (ESCAPE) twice, to return to
measurement mode.