8.7 Viewing TDS Calibration Data – P5.0
Program 5 is a “view only” option, which displays the value at which calibration was
performed.

P5.1: view Calibrated TDS value (if no calibration performed for this range, controller
displays ‘- - - - ‘.

These parameters will change each time you recalibrate the controller.

P5.0

C L

P5.1

CAL

300

Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and return
to Measurement mode.

P5.0: Viewing Calibrated TDS data
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P5.0 and CAL. Press ENTER.

3.

Screen will scroll P5.1, ‘CAL’, then Calibrated value (in ppm or ppt).

4.

Press ▲and ▼keys together (ESCAPE) twice, to return to measurement mode.