6 technical specifications – Thermo Fisher Scientific Alpha pH 600 RF (In the process of updating the brand guideline & product image) User Manual
Page 38
- 32 -
6
T E C H NI C AL S PE C IFI C AT IO N S
General Specification
(a) pH
pH Measuring Range
0.00 to 14.00 pH
Resolution 0.01
pH
Accuracy
± 0.01 pH
(b) mV
mV Measuring Range
-1000 to 1000 mV
Resolution 1
mV
Accuracy
± 1 mV
(c) Temperature
Temperature measuring range
-10.0 to +110.0
o
C or +14.0 to +230.0
o
F
Resolution 0.1
o
C or 0.1
o
F
Relative accuracy
0.5
o
C or 1.0
o
F
Sensor
3 wire / 2 wire – Pt100 or Pt1000 (Jumper Selectable)
Compensation
Automatic or Manual
Calibration Point
(a) pH
Number of calibration points
Number of calibration buffers
1 or 2 points
USA: 4.01, 7.00, 10.01
NIST: 4.01, 6.86, 9.18
(c) Temperature
Offset Adjustment
±10 ˚C/ ±18 ˚F
Input / Output
Input
Asymmetrical / Symmetrical
Output
UART Serial Comm. Output for the RF RX/TX Board:
Protocol : 8-N-1
Data rate: 9,600 kbps
Display
LCD
UV coat, Custom Dual segments display with symbols
for Status information.
Back light
On/Off selectable
Electrical data and Connections
pH/mV Input
8-pin Screw Terminal (3.5mm pitch)
Temperature Input
8-pin Screw Terminal (3.5mm pitch)
Power
Phone Jack DC Socket
RF Communication Specifications
Transmission Method
TDMA, FSK Modulation
Frequency
923 to 928 MHZ, ISM Band Region 2
Air Baud Rate
16 Kbit/S
Range Single Unit
100 m (328 ft) (in open air)
RF Power
+5 dB
Antenna
Internal diversity antenna
Network Model
Mesh
Operating Temperature
-10 to 65 °C
Others
Power Input
+9 V DC
Pollution Degree
2
Dimensions (W x H x D)
96mm x 96mm x 66 mm
Weight (Estimated)
210g
Ambient Temp. operating range
0 to 40
o
C
Maximum Relative Humidity
80% up to 31
o
C decreasing linearly to 50% at 40
o
C
Ingress Protection
IP65