5.6

Viewing Electrode Properties

Each time you calibrate your pH electrode, the pH Transmitter re-calculates slope &
offset of the electrode and shown in the LCD at the end of calibration. The setup
mode allows you to view slope & offset values at any time.

NOTE:

When the measurement mode is set to ORP, only the offset value of
the electrode is displayed.

SETUP

HOLD

mV

2

- 31 -

pH

SETUP

ENT

1

ENT

mV

2

measurement mode

measurement mode

SETUP

For pH

For ORP

ENT

11

To view the calibration and parameter, press CAL key, then ENT key.

2

Electrode status: T

de annunciator

he electro

appears in the LCD. For

pH measurement mode, the slope (in mV) is shown in the upper display
and normalized temperature is shown in the lower display.
For ORP measurement mode, the offset value (in mV) of the electrode is
shown in the upper display.
Press ENT key to quit from electrode properties screen.

NOTE:

To exit from any intermediate steps, press ▲ and ▼ keys (Escape)
together. It will return to measurement mode.