5.6
Viewing Electrode Properties
Each time you calibrate your pH electrode, the pH Transmitter re-calculates slope &
offset of the electrode and shown in the LCD at the end of calibration. The setup
mode allows you to view slope & offset values at any time.
NOTE:
When the measurement mode is set to ORP, only the offset value of
the electrode is displayed.
SETUP
HOLD
mV
2
- 31 -
pH
SETUP
ENT
1
ENT
mV
2
measurement mode
measurement mode
SETUP
For pH
For ORP
ENT
11
To view the calibration and parameter, press CAL key, then ENT key.
2
Electrode status: T
de annunciator
he electro
appears in the LCD. For
pH measurement mode, the slope (in mV) is shown in the upper display
and normalized temperature is shown in the lower display.
For ORP measurement mode, the offset value (in mV) of the electrode is
shown in the upper display.
Press ENT key to quit from electrode properties screen.
NOTE:
To exit from any intermediate steps, press ▲ and ▼ keys (Escape)
together. It will return to measurement mode.