4.4

Check Calibration

MEAS

pH

READY

ATC

pH

CAL

pH

mV

CAL

ENT

ENT

1. While in Measurement mode press the CAL key. The display will prompt you to

enter a security code. Leave the security code at “0” (view only mode).

2. Press

the

ENT key.

The display shows the slope in mV (Main Reading) and the normalized
temperature is shown on the Auxiliary reading of the display.

NOTE:

To exit from any intermediate steps, press ▲ and ▼ keys (Escape)
together. It will return to measurement mode.

4.5

ORP Calibration

If the unit of measurement is in mV, then the calibration is also carried out in mV.

CAL

mV

HOLD

CAL

m V

HOLD

CAL

m V

HOLD

m V

ENT

CAL

°C

READY

1. Enter Calibration mode as described in section 4.2. The display shows “CAL

ORP” (for mV calibration) and the “CAL” indicator appears on the display.

2. Place sensor in the solution.

3. Press

the

ENT key to start calibration. The main display shows the current mV

output of the electrode with offset adjustment and auxiliary display shows the
mV output with out any offset adjustments.

4. Press

the

▲ or ▼ key to match the mV value to exact reading.

- 21 -