- 17 -

11

From pH measurement mode press CAL key to enter calibration mode
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to
begin calibration.

2

Place the electrode in pH 7.00 buffer. Immerse the temperature probe in
the buffer solution if ATC mode is enabled. Immerse the potential
matching pin in the buffer if symmetrical mode is enabled. The buffer

annunciator

appears in LCD. Lower display shows pH 7.00 (USA

standard buffer). Upper display shows the current uncalibrated pH
reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator
when the reading is stable. Press CAL key to confirm the reading.

3

The calibration is completed. The pH Transmitter re-calculates electrode
properties based on the calibration. The new slope (in mV) is shown in the
upper display and normalized temperature is shown in lower display.
Press ENT key to exit from the calibration.

4

The pH Transmitter reverts to pH measurement mode.

NOTE:

Refer notes at the end of this section for additional information and
possible error indicators of calibration process

2-Point Calibration:

CAL

CAL

pH

READY

pH

CAL

pH

READY

pH

CAL

pH

mV

MEAS

READY

CAL

pH

READY

pH

ENT

ENT

ENT

ENT

ENT

CAL

°C

ATC

pH

1

1

1

2

3

4

5

5

3

1

From pH measurement mode press CAL key to enter calibration mode
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to
begin first calibration point.

2

Place the electrode in pH 7.00 buffer. Immerse the temperature probe in
the buffer solution if ATC mode is enabled. Immerse the potential
matching pin in the buffer if symmetrical mode is enabled. The buffer

annunciator

appears in LCD. Lower display shows pH 7.00 (USA

standard buffer). Upper display shows the current uncalibrated pH
reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator